Wired CR Radchex Meter
Wired CR Radchex Meter
Technical Specifications
X-ray energy dependence: Simulates relative light output of Photostimulatable Phosphor Plate (PSP) within +/- 3% over kVp range of 60 to 120 kVp and a patient equivalent thickness range of 5 to 35 cm (within specified operating rates).

Digital range
: Computed Radiography Light Units; CRLU (AEC#); 0 to 655.35

Wired CR Radchex Minimum CRLU Rate
: 1.50/sec (approx 0.15 mR/sec entrance exposure rate)

Maximum CRLU Rate
: 2500/sec (approx 250 mR/sec entrance exposure rate)

Power On/Off: Manual switch

Controls: Manually using the Power/Reset button.

Functions: Measures CRLU (AEC#); Converts CRLU to CR manufacturers specific CR plate reader light exposure index value (EI); user selectable. Calculates cassette input exposure values for various X-ray beam conditions (exposure in mR plus backscatter)

Power requirements: 2 x 9Volt Batteries in Interface box.

Typical battery life: Greater than 8 hrs.

Operating environment
: 59 degrees to 95 degrees F (15 to 35 degrees C)

X-ray beam filter: 1.5 mm copper (B152-110); 6 inches by 6 inches complete with velcro straps to attach to X-ray tube collimator housing

Electronic Cassette
:
Dimensions: 10 x 12 x 0.5 inches (24 x 30 x 1.3 cm)
Weight: 3.9 lb (1.8 kg)

Software
: Microsoft Excel Template and Instruction Manual on included CD.
Download Documentation in .pdf format.
 